CHAPTER 7: Image Management-ViewStation & Photos and Diagrams<br />

7. Click Close twice to close the Treatment Delivery Table dialog box and Treatment<br />

Chart window.<br />

CAUTION<br />

Planar Image Registration<br />

If you are treating a field, when you click Select in the Treatment<br />

Delivery Table dialog box, the Port Film Status window opens if the<br />

field has associated images that require changes or reviews or has a<br />

Voided or Rejected review status.<br />

The system only issues the alert to open the Port Film Status<br />

window if the Portal Image Check check box is selected in the<br />

system utilities. Click File | System Utilities | Department<br />

Setup…, then click the Clinical (RO) tab to make sure that this<br />

check box is selected. If the Portal Image Check check box is not<br />

selected, contact your system administrator.<br />

Image Registration is the process that you use to fuse a verification (portal/treatment) image<br />

with a reference (DRR or simulator) image then calculate positional offsets for patient<br />

positioning. Given the two images with their corresponding beam isocenters, <strong>MOSAIQ</strong><br />

computes the translational shifts required to align the patient to match the anatomy in the<br />

verification image with the anatomy in the reference image. The difference between the beam<br />

isocenters of these images yields the computed positional shifts. (The computed positional<br />

shifts are shifts required to move the patient so that he/she is properly aligned with respect to<br />

the planned treatment setup.) The verification (portal) image lets you review current beam<br />

position relative to intended position as shown on the reference (DRR or simulator) image.<br />

CAUTION<br />

CAUTION<br />

In the registration process, make sure that the reference and<br />

treatment images are in Beam's Eye View coordinates. That is, the<br />

anatomy in the images appears as if the gantry axis is<br />

perpendicular to the top of the image viewer and the gantry stand is<br />

at the top of the window (gantry-up position). Use applicable caution<br />

when you flip images to be used by the registration process. (If you<br />

flip an image, it does not have the same effect as if you rotated the<br />

same image 180˚.) It is your responsibility to review calculated shift<br />

direction and magnitude before you save images. We recommend<br />

that you make modifications if clinically necessary.<br />

You must have Image | ViewStation | Review Images | Enable and<br />

Image | ViewStation | Image Information | Modify security rights to<br />

register treatment images for planar (2D) shift analysis.<br />

Copyright 2010, Elekta Impac Software 191<br />

LUGMSQ0002/4.0/23-Sep-2010