Localization Trend Review<br />

CHAPTER 7: Image Management-ViewStation & Photos and Diagrams<br />

Localization Trend Review lets you examine the patient setup correction and session offset<br />

history. When you complete an image registration and click OK in the Image Review<br />

window, the Localization Trend Review dialog box opens. The Localization Trend Review<br />

dialog box shows active patient offset data. You can quickly examine the history of recorded<br />

setup corrections for past treatment sessions and not open a series of images. You can open<br />

the Localization Trend Review dialog box from the Treatment Delivery table, Diagnoses and<br />

Interventions window, and the Localization Review Worklist.<br />

Interfraction setup variation, as measured with setup imaging, consists of random and<br />

systematic components. You can examine Session Offsets for patterns and correct systematic<br />

setup variation during setup prior to verification imaging. When you remove systematic error<br />

from treatment setup, the daily workflow efficiency increases and may reduce dependency on<br />

daily setup imaging for many treatment sites. The systematic component of a treatment<br />

course is equal to the average setup variation over a course of treatment. You can closely<br />

determine this variation with measurements from the first several treatment sessions.<br />

Setup Intelligence refers to systematic variation correction as a Localization Offset. You<br />

include only the initial Session Offset for a series of treatment sessions when you calculate a<br />

Localization Offset. The constant is the isocenter position relative to the localization<br />

references (tattoos). You should assign any offsets you record when you verify a new couch<br />

position after positional verification a status of Inactive or Excluded.<br />

The Localization Trend Review dialog box opens, and shows offsets with an Active status<br />

only. The Localization Trend Review Report lets you select filters such as Show History. The<br />

report can show the current and/or historic offset information. Refer to Figure 59.<br />

Figure 59: Localization Trend Review Dialog Box<br />

Copyright 2010, Elekta Impac Software 213<br />

LUGMSQ0002/4.0/23-Sep-2010