<strong>MOSAIQ</strong> User’s <strong>Guide</strong><br />

SEQUENCER Workflows<br />

SEQUENCER connects to your linac interactively or statically, depending on the interface.<br />

An interactive interface lets SEQUENCER communicate directly with the linac in real time.<br />

SEQUENCER must wait for data transfers when the interface is static. An example of an<br />

interactive interface is the interface to an Elekta Synergy linac. An example of a static<br />

interface is the interface to a Siemens Coherence linac. SEQUENCER sends the treatment<br />

field to the linac and verifies that the linac is correctly moded up before you can beam-on.<br />

The Varian 4DITC and Siemens Syngo is a special case because the interface only lets<br />

SEQUENCER record geometric machine and treatment data.<br />

Basic SEQUENCER Workflow<br />

The steps below apply to any SEQUENCER interface configuration after you select the<br />

patient and open the patient chart. Refer to Select a Patient for Treatment on page 261 in<br />

this section.<br />

1. Select the Treatment Calendar session that you want.<br />

SEQUENCER sends the treatment field data to the linac and opens the Verified<br />

Treatment window. You can use the Auto Setup option to setup the linac to deliver the<br />

selected treatment field.<br />

2. Select the first field to treat, then click Select<br />

If the Treatment Readiness Check dialog box opens, you must respond before you can<br />

continue.<br />

3. Setup the patient on the couch.<br />

If your department uses accessory barcodes, you must scan the accessory barcodes now<br />

to avoid an interlock.<br />

4. Make pre-treatment portal images if necessary.<br />

5. Mode up the linac.<br />

SEQUENCER compares the prescribed parameter values against the machine parameter<br />

values. SEQUENCER prevents treatment if one or more parameters is out of tolerance.<br />

6. Beam-on the linac.<br />

***Treating*** appears below the header. The Record Treatment dialog box opens<br />

when the treatment is complete.<br />

7. Treat the first field and make during and or post-treatment portal images if necessary.<br />

8. Type any relevant comments in the Note field of the Record Treatment dialog box.<br />

9. Click Record.<br />

The Verified Treatment window opens for the next field because you<br />

set the treatment order earlier. If you had not set the treatment<br />

sequence, the Treatment Delivery Table window will open after you<br />

record the treatment.<br />