<strong>MOSAIQ</strong> User’s <strong>Guide</strong><br />

1. Follow the instructions in the Treatment Readiness Check dialog box.<br />

If the highlight color is yellow or green, click Continue. If the highlight color is red, you<br />

must override or change any field for which SEQUENCER prevented treatment.<br />

2. Click Continue after you override.<br />

If your linac is a Varian 4DITC or Siemens Syngo, SEQUENCER makes an outbound check<br />

after you click Send Plan from the Treatment Delivery Table. This is to make sure that any<br />

changes made (when you change or hide fields, etc.) do not cause additional issues.<br />

Field Sequencing Verification Checks<br />

The Treatment Readiness Check does the same validations as the Treatment Calendar. If<br />

<strong>MOSAIQ</strong> finds errors in any field in the CFS (Consolidated Field Sequencing) during the<br />

standard Treatment Readiness Check treatment delivery with either a CFS or an AFS<br />

(Automatic Field Sequencing) session is prevented. You must treat the fields on the<br />

Treatment Calendar separately. You must acknowledge warnings or override parameters as<br />

noted by the Treatment Readiness Check to deliver treatment. You can manually stop the<br />

CFS session to work around unresolved errors requiring acknowledgements or overrides.<br />

A valid CFS must include the following:<br />

All fields must have the same treatment machine.<br />

All fields must be of Type Static or StepNShoot.<br />

All fields must be of Modality X-Rays.<br />

All fields must have an MLC defined.<br />

No fields may have a physical wedge.<br />

No fields may have the text “CUSTOM” for compensator, block, or bolus.<br />

All fields must have the same defined text in the block code, if a block code is<br />

present in any of the fields.<br />

All fields must have the same defined text in the bolus code, if a bolus code is present<br />

in any of the fields.<br />

All fields must have the same defined text in the compensator code, if a compensator<br />

code is present in any of the fields.<br />

All fields must have the same couch positions or be within +/- 1mm tolerance of the<br />

first field in the session.<br />

The CFS sequence may pass initial Treatment Calendar CFS checks, but one or more fields<br />

within the CFS session may not pass the additional treatment readiness checks. In that case,<br />

you can either resolve an outstanding error during the session or skip those unresolved errors<br />

in order to treat the eligible fields separately.<br />

Port Film Status<br />

The Port Film Status window shows the results of the pre-treatment portal image check.<br />

1. Click View to open the Image Review window.<br />

2. Click OK or Cancel to open the Verified Treatment window.<br />

262 Copyright 2010, Elekta Impac Software<br />