CHAPTER 6: MLC Fit<br />

The display may be affected by the current image state: If you change the gantry angle, couch<br />

angle, or couch top using edit controls, capture settings, or import, the Reference Image<br />

becomes unavailable.<br />

Geometry display in Image: Check to show the treatment geometry on the image, and uncheck<br />

to hide the treatment geometry layer. This check box is only available when the<br />

following ten conditions occur:<br />

The image is for the currently selected patient.<br />

The image is a Simulator or DRR image.<br />

The image is not an image series.<br />

The image is enabled for Tx Definition.<br />

The image is a reference image and is reticle-scaled and has square pixels.<br />

The image is associated with the currently selected field, and the treatment machine<br />

assigned to the field is characterized. The treatment geometry comes from the<br />

associated treatment field definition.<br />

The Shadow filter is not applied.<br />

The image does not have an incomplete Re-Image change request.<br />

The image is not voided.<br />

The image is not archived.<br />

<strong>MOSAIQ</strong> alerts you when the image is archived.<br />

<strong>MOSAIQ</strong> alerts you when you manually display reference image for the first time because it<br />

did not automatically appear.<br />

<strong>MOSAIQ</strong> alerts you when another user edits the image between the time you opened the<br />

Beam’s Eye View dialog box and attempt to save, aborting the save and keeping the Beam’s<br />

Eye View dialog box open. You can only close the dialog box by canceling.<br />

You can change the Vertical, Lateral, and Longitudinal couch<br />

parameters, and the reference image remains available.<br />

Beam’s Eye View Calibration Point Digitization<br />

The Beam’s Eye View Calibration Point Digitization dialog box prompts you to digitize a<br />

point and gives the coordinates of the point digitized. After digitizing each point, a “beep”<br />

sounds and you are prompted to enter the next point. Upon digitizing the second point the<br />

transformation parameters are computed. After digitizing the third and fourth points, the<br />

scaled digitized values are compared with the specified values. If the scaled value is not<br />

within the specified tolerance of the desired value, the calibration process terminates.<br />

Beam’s Eye View Shape Rotation<br />

This dialog box enables you to enter a new angle to rotate the treatment shape. If the<br />

orientation of the reticle angle was incorrect when you digitized the treatment shape, you can<br />

simply rotate the shape using this dialog box instead of digitizing a new shape.<br />

Copyright 2010, Elekta Impac Software 135<br />

LUGMSQ0002/4.0/23-Sep-2010