Figure 53: Planes of a Stereoscopic Pair of Images<br />

The following instructions provide one example of the workflow when registering images at<br />

the linac with Setup Intelligence.<br />

1. Right-click or press F4 on the Treatment Delivery Table, then click Image Review… to<br />

open the Image Review window.<br />

2. Click the Isocenter Offset tab, then click Register to register a stereoscopic pair of<br />

images.<br />

<strong>MOSAIQ</strong> automatically looks for any potential stereoscopic pair images on the work<br />

list. These images must meet gantry separation requirements and be taken within 10<br />

minutes of the original image. For example, when viewing the PA image and a lateral or<br />

oblique image was taken within 10 minutes, you have the option to apply a stereoscopic<br />

registration.<br />

When more than one image meets these criteria, the Stereoscopic Image Selection dialog<br />

box opens. Refer to Figure 54.<br />

Figure 54: Stereoscopic Image Selection Dialog Box<br />

208 Copyright 2010, Elekta Impac Software<br />

LUGMSQ0002/4.0/23-Sep-2010