Air quality expert group - Fine particulate matter (PM2.5) in ... - Defra
Air quality expert group - Fine particulate matter (PM2.5) in ... - Defra
Air quality expert group - Fine particulate matter (PM2.5) in ... - Defra
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>PM2.5</strong> <strong>in</strong> the UK<br />
36<br />
57. Black carbon data have the advantages of sensitivity, high time resolution<br />
and high data capture, because of the relatively simple physical pr<strong>in</strong>ciples of<br />
the <strong>in</strong>strument. However, <strong>in</strong>terpretation of the data should take <strong>in</strong>to account<br />
the fact that the measurement does not correspond to a specific chemical<br />
component of the PM. The measured parameter is similar to elemental<br />
carbon, but this metric itself is not precisely def<strong>in</strong>ed, as described below. Also,<br />
standardisation and QA/QC procedures for black carbon are at a less advanced<br />
stage than for regulated pollutants, so that comparisons with data from outside<br />
the Network will have limitations.<br />
2.4.2 PM10 nitrate, sulphate and chloride<br />
58. Daily measurements of the <strong>in</strong>organic components of PM10 (sulphate, nitrate and<br />
chloride) are made with<strong>in</strong> the Particle Numbers and Concentrations Research<br />
Network, 3 which uses a Thermo Partisol 2025 sequential air sampler. S<strong>in</strong>ce<br />
2007, ultrapure quartz filters (Pallflex Tissuquartz) have been used and cut <strong>in</strong><br />
half to allow for the analysis of both elemental carbon/organic carbon and<br />
<strong>in</strong>organic components. Further details about the Network are available at<br />
http://uk-air.defra.gov.uk.<br />
59. The Partisol 2025 sampler, as described <strong>in</strong> Section 2.3.3, provides un<strong>in</strong>terrupted<br />
sampl<strong>in</strong>g of ambient air and automatic exchange of filters for up to 16 days.<br />
The <strong>in</strong>strument uses an airflow of 1 m 3 hr -1 through a PM10 <strong>in</strong>let and the<br />
filter temperature is ma<strong>in</strong>ta<strong>in</strong>ed to with<strong>in</strong> ±5°C of ambient temperature. The<br />
exposed filters are stored <strong>in</strong> small polypropylene filter bags and kept <strong>in</strong> a cold<br />
room until analysis to prevent further loss of semi-volatile components such as<br />
ammonium nitrate. Extracts from the filters are dissolved <strong>in</strong> an eluent of 3.5 mM<br />
sodium carbonate and 1 mM sodium hydrogen carbonate and analysed <strong>in</strong> the<br />
laboratory by ion chromatography for sulphate, nitrate and chloride content.<br />
Ambient concentrations are derived from the mass measured on the filter<br />
and the airflow dur<strong>in</strong>g the sampl<strong>in</strong>g period. Calibration is via gravimetricallyprepared<br />
solutions.<br />
60. CEN has produced a Technical Report standardis<strong>in</strong>g these measurements, TR<br />
16269:2011 Ambient air: Guide for the measurement of anions and cations<br />
<strong>in</strong> <strong>PM2.5</strong>. In practice, although standardisation is less advanced than for other<br />
regulated pollutants, the analytical process is relatively simple, and analytical<br />
results should be relatively robust. Comparisons with other data are likely to be<br />
more compromised by losses of semi-volatile material before analysis, especially<br />
of ammonium nitrate, which can be a substantial fraction of the nitrate and the<br />
<strong>PM2.5</strong>, as described <strong>in</strong> Chapter 3.<br />
2.4.3 PM10 elemental carbon/organic carbon<br />
61. Daily measurements of the elemental carbon/organic carbon (EC/OC)<br />
components of PM10 are made with<strong>in</strong> the Particle Numbers and Concentrations<br />
Research Network. A Thermo Partisol 2025 sequential air sampler is used, as for<br />
the sulphate, nitrate and chloride components above. Ultrapure quartz filters<br />
(Pallflex Tissuquartz) are used.<br />
3 Operated by the National Physical Laboratory on behalf of <strong>Defra</strong> and the Devolved Adm<strong>in</strong>istrations.