Using <strong>the</strong> <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> 5<br />

<strong>Capillary</strong> <strong>Electrophoresis</strong>, CE<br />

Table 24<br />

Alignment interfaces<br />

Sleeve Color Part Number Used For *<br />

Black G<strong>7100</strong>-60150 25 µm id capillaries with extended light path<br />

Green G<strong>7100</strong>-60210 50 µm id straight capillaries<br />

Red G<strong>7100</strong>-60230 50 µm id capillaries with extended light path<br />

Blue* G<strong>7100</strong>-60310 75 µm id straight capillaries<br />

Yellow G<strong>7100</strong>-60330 75 µm id capillaries with extended light path<br />

* Use <strong>the</strong> blue alignment interface for 100 or 150 µm id capillaries<br />

1 Make sure <strong>the</strong> color of <strong>the</strong> sleeve in <strong>the</strong> alignment interface matches <strong>the</strong><br />

color of <strong>the</strong> plastic s<strong>to</strong>p on <strong>the</strong> capillary.<br />

2 Remove <strong>the</strong> protective plastic cap that covers <strong>the</strong> detection window.<br />

NOTE<br />

The capillary is very fragile at that point and it is important <strong>to</strong> keep <strong>the</strong> detection<br />

window clean.<br />

• Do not <strong>to</strong>uch <strong>the</strong> capillary at <strong>the</strong> detection window or <strong>the</strong> capillary tips, where <strong>the</strong><br />

polyimide is burnt off.<br />

• Clean <strong>the</strong> detection window and <strong>the</strong> capillary tips as described. Put <strong>the</strong> insertion<br />

<strong>to</strong>ol on a flat surface, for example, your lab bench.<br />

Colored sleeve<br />

Alignment interface<br />

<strong>Capillary</strong> insertion <strong>to</strong>ol<br />

Figure 28<br />

Alignment interface and insertion <strong>to</strong>ol<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 125