LAN Configuration 4<br />

Initialization Mode Selection<br />

Using S<strong>to</strong>red<br />

When initialization mode "Using S<strong>to</strong>red" is selected, <strong>the</strong> parameters are taken<br />

from <strong>the</strong> non-volatile memory of <strong>the</strong> instrument. The TCP/IP connection will<br />

be established using <strong>the</strong>se parameters. The parameters were configured<br />

previously by one of <strong>the</strong> described methods.<br />

Figure 5<br />

Using S<strong>to</strong>red (principle)<br />

Using Default<br />

When “Using Default” is selected, <strong>the</strong> fac<strong>to</strong>ry default parameters are used<br />

instead. These parameters enable a TCP/IP connection <strong>to</strong> <strong>the</strong> LAN interface<br />

without fur<strong>the</strong>r configuration, see Table 19 on page 68.<br />

A direct connection of a computer and <strong>the</strong> <strong>7100</strong> instrument is <strong>the</strong> entry point<br />

<strong>to</strong> change <strong>the</strong> s<strong>to</strong>red parameters of <strong>the</strong> instrument. The default parameters of<br />

<strong>the</strong> <strong>7100</strong> instrument are permanently s<strong>to</strong>red and cannot be changed. Using<br />

<strong>the</strong>m, this allows you <strong>to</strong> establish a LAN connection at any time when <strong>the</strong> init<br />

mode “Using Default” is configured.<br />

Figure 6<br />

Using Default (principle)<br />

NOTE<br />

Using <strong>the</strong> default address in your local area network may result in network problems. Take<br />

care and change it <strong>to</strong> a valid address immediately unless you plan a direct LAN connection<br />

<strong>to</strong> stand-alone PC.<br />