7 Hardware Information<br />

External Water Bath for Tray Cooling (Optional)<br />

External Water Bath for Tray Cooling (Optional)<br />

The two connections on <strong>the</strong> left rear of <strong>the</strong> instrument (see Figure 2 on<br />

page 59) can be used <strong>to</strong> install an external water bath for <strong>the</strong> tray of <strong>the</strong><br />

<strong>Agilent</strong> CE instrument.<br />

The temperature for <strong>the</strong> cooling liquid is set at <strong>the</strong> external water bath. The<br />

temperature display in <strong>the</strong> CE diagram displays <strong>the</strong> actual temperature of <strong>the</strong><br />

tray controlled by <strong>the</strong> cooling liquid. The external water bath cannot be<br />

controlled by <strong>the</strong> <strong>Agilent</strong> CE instrument.<br />

Good Practices<br />

To avoid problems when using <strong>the</strong> external water bath, consider <strong>the</strong> following<br />

points.<br />

• The tubes used should have <strong>the</strong>rmal insulation <strong>to</strong> avoid condensation.<br />

• Use a displacer <strong>to</strong> minimize <strong>the</strong> amount of liquid that is released in case of<br />

leaking cooling lines.<br />

• The device should be equipped with a liquid level sensor <strong>to</strong> switch off<br />

au<strong>to</strong>matically if liquid is lost.<br />

• Do not operate <strong>the</strong> devices at temperatures below +1 °C <strong>to</strong> avoid formation<br />

of ice on internal surfaces.<br />

External Water Bath Requirements<br />

• Minimum cooling power 100 W, at bath temperatures of 5 °C.<br />

• Temperature precision better than ± 0.2 °C.<br />

• Safety requirements must meet national standards and also: be suitable for<br />

unattended operation, be suitable for continuous operation, and have<br />

controllable high temperature protection.<br />

• Requirements for external liquid circulation:<br />

Suitable for external circulating of liquid,<br />

Suitable for connection of 8 mm ID tubes,<br />

Minimum circulating liquid flow rate 8 L/min,<br />

184 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual