Introduction to the Agilent 7100 Capillary Electrophoresis System
Introduction to the Agilent 7100 Capillary Electrophoresis System
Introduction to the Agilent 7100 Capillary Electrophoresis System
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
7 Hardware Information<br />
External Water Bath for Tray Cooling (Optional)<br />
External Water Bath for Tray Cooling (Optional)<br />
The two connections on <strong>the</strong> left rear of <strong>the</strong> instrument (see Figure 2 on<br />
page 59) can be used <strong>to</strong> install an external water bath for <strong>the</strong> tray of <strong>the</strong><br />
<strong>Agilent</strong> CE instrument.<br />
The temperature for <strong>the</strong> cooling liquid is set at <strong>the</strong> external water bath. The<br />
temperature display in <strong>the</strong> CE diagram displays <strong>the</strong> actual temperature of <strong>the</strong><br />
tray controlled by <strong>the</strong> cooling liquid. The external water bath cannot be<br />
controlled by <strong>the</strong> <strong>Agilent</strong> CE instrument.<br />
Good Practices<br />
To avoid problems when using <strong>the</strong> external water bath, consider <strong>the</strong> following<br />
points.<br />
• The tubes used should have <strong>the</strong>rmal insulation <strong>to</strong> avoid condensation.<br />
• Use a displacer <strong>to</strong> minimize <strong>the</strong> amount of liquid that is released in case of<br />
leaking cooling lines.<br />
• The device should be equipped with a liquid level sensor <strong>to</strong> switch off<br />
au<strong>to</strong>matically if liquid is lost.<br />
• Do not operate <strong>the</strong> devices at temperatures below +1 °C <strong>to</strong> avoid formation<br />
of ice on internal surfaces.<br />
External Water Bath Requirements<br />
• Minimum cooling power 100 W, at bath temperatures of 5 °C.<br />
• Temperature precision better than ± 0.2 °C.<br />
• Safety requirements must meet national standards and also: be suitable for<br />
unattended operation, be suitable for continuous operation, and have<br />
controllable high temperature protection.<br />
• Requirements for external liquid circulation:<br />
Suitable for external circulating of liquid,<br />
Suitable for connection of 8 mm ID tubes,<br />
Minimum circulating liquid flow rate 8 L/min,<br />
184 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual