Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

Accessing <strong>the</strong> Electrodes<br />

1 Use a screwdriver <strong>to</strong> unscrew <strong>the</strong> two metal screws holding <strong>the</strong> insulation<br />

plate.<br />

2 Use a screwdriver <strong>to</strong> remove <strong>the</strong> screw that secures <strong>the</strong> blue ground cable.<br />

Metal screws<br />

Insulation plate<br />

Figure 58<br />

Insulation plate<br />

3 Use <strong>the</strong> 12 mm hexagonal socket screwdriver <strong>to</strong> unscrew <strong>the</strong> inlet electrode.<br />

12 mm hexagonal<br />

socket screwdriver<br />

High voltage cable, red<br />

Ground cable screw<br />

Ground cable, blue<br />

Figure 59<br />

Removing <strong>the</strong> ground cable and <strong>the</strong> inlet electrode cable<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 199