Maintenance 8<br />

Changing <strong>the</strong> Lamp<br />

Removing <strong>the</strong> Existing Lamp<br />

Parts required<br />

Pozidriv screwdriver (part of <strong>the</strong> accessory kit).<br />

Deuterium lamp for <strong>the</strong> <strong>Agilent</strong> CE instrument.<br />

1 Select Exit from <strong>the</strong> File menu <strong>to</strong> exit <strong>the</strong> <strong>Agilent</strong> ChemStation. Exit both<br />

online and offline copies.<br />

2 Turn off <strong>the</strong> line power <strong>to</strong> <strong>the</strong> <strong>Agilent</strong> CE instrument.<br />

3 Open <strong>the</strong> <strong>to</strong>p cover.<br />

4 Open <strong>the</strong> detec<strong>to</strong>r cover.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 225