8 Maintenance<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

Buffer precipitates, dust, moisture or o<strong>the</strong>r solid material in <strong>the</strong> electrodes<br />

and pre-punchers or under <strong>the</strong> insulation plate may cause arcing. They can<br />

lead <strong>to</strong> current leakage, buffer contamination or carry-over. Cleaning <strong>the</strong>se<br />

parts regularly will prevent this.<br />

The electrodes, pre-punchers and <strong>the</strong> insulation plate should be cleaned<br />

regularly. Depending on <strong>the</strong> application, cleaning frequency can be for<br />

example once a week, or if peaks show unusual tailing or unusual peak<br />

shoulders. Please be aware that <strong>the</strong> EMF counter analysis start lets you set<br />

cus<strong>to</strong>mized limits for warnings. Cleaning is necessary if arcing or excessive<br />

current leakage is observed. Usage of creeping buffers, such as urea, may lead<br />

<strong>to</strong> higher cleaning frequency.<br />

Steps for <strong>the</strong> cleaning procedure for electrodes, pre-punchers and <strong>the</strong><br />

insulation plate:<br />

• Prepare <strong>the</strong> <strong>Agilent</strong> CE instrument for this task<br />

• Remove <strong>the</strong> covers<br />

• Access <strong>the</strong> electrodes<br />

• Access <strong>the</strong> liquid handling module<br />

• Access <strong>the</strong> pre-punchers<br />

• Clean <strong>the</strong> electrodes<br />

• Clean <strong>the</strong> pre-punchers<br />

• Clean <strong>the</strong> insulation plate<br />

• Reinstall <strong>the</strong> pre-punchers<br />

• Reinstall <strong>the</strong> insulation plate<br />

• Reinstall <strong>the</strong> electrodes<br />

NOTE<br />

Follow <strong>the</strong> common safety regulations for labora<strong>to</strong>ries when doing this task. Wear safety<br />

glasses and rubber gloves.<br />