LAN Configuration 4<br />

Initialization Mode Selection<br />

Initialization Mode Selection<br />

The following initialization (init) modes are selectable:<br />

Table 18<br />

Initialization mode switches<br />

SW6 SW7 SW8 Init mode<br />

OFF OFF OFF Bootp<br />

OFF OFF ON Bootp & S<strong>to</strong>re<br />

OFF ON OFF Using S<strong>to</strong>red<br />

OFF ON ON Using Default<br />

Bootp<br />

When <strong>the</strong> initialization mode "Bootp" is selected, <strong>the</strong> <strong>7100</strong> CE tries <strong>to</strong><br />

download <strong>the</strong> parameters from a Bootp server. The parameters obtained<br />

become <strong>the</strong> active parameters immediately. They are not s<strong>to</strong>red in <strong>the</strong><br />

non-volatile memory of <strong>the</strong> <strong>7100</strong> CE. Therefore, <strong>the</strong> parameters are lost with<br />

<strong>the</strong> next power cycle of <strong>the</strong> <strong>7100</strong> CE.<br />

Figure 3<br />

Bootp (principle)<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 65