5 Using <strong>the</strong> <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong><br />

<strong>Capillary</strong> <strong>Electrophoresis</strong>, CE<br />

Vials<br />

You can program a method <strong>to</strong> use certain vials as <strong>the</strong> inlet or outlet home vial.<br />

During electrophoresis, <strong>the</strong> electrodes and <strong>the</strong> capillary are in <strong>the</strong>se vials. Vial<br />

numbers 1 <strong>to</strong> 48 or user vials (User 1 <strong>to</strong> User 10) can be selected. The selection<br />

can be done from <strong>the</strong> pull-down menu or by typing in <strong>the</strong> number with <strong>the</strong><br />

keyboard. Clicking Add Method Vial allows you <strong>to</strong> enter vial names which will be<br />

visible in <strong>the</strong> vialtable.<br />

Figure 22 Add Method Vial<br />

User vials are numbered from 1 <strong>to</strong> 10. User vials are placeholders in <strong>the</strong><br />

method which are resolved by <strong>the</strong> sequence table at a later point in time.<br />

Usage of user vials is helpful for method optimization. Various buffer<br />

compositions can be tested with one common method. Creation of methods<br />

with variations of Inlet or outlet home vials can be omitted. Usage of user vials<br />

is also helpful <strong>to</strong> change buffers in long sequences <strong>to</strong> avoid over utilization of<br />

<strong>the</strong> buffer.<br />

NOTE<br />

Vial numbers below 10 must be typed in with a preceding 0, for example 02 for a vial in tray<br />

location number 2, since <strong>the</strong> vials are numbered with two digits.<br />