Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

10 Tighten <strong>the</strong> slotted screws in <strong>the</strong> front of <strong>the</strong> liquid handling module.<br />

3. Front screws of <strong>the</strong><br />

liquid handling module<br />

2. Put in bottles<br />

1. Push liquid handling<br />

module in<br />

Figure 70<br />

Liquid handling module<br />

11 Reconnect <strong>the</strong> line power, drainage, external pressure supply, <strong>the</strong><br />

connec<strong>to</strong>rs for an external cooling bath (if applicable), and switch on <strong>the</strong><br />

instrument.<br />

12 Reestablish <strong>the</strong> connection <strong>to</strong> <strong>the</strong> instrument from <strong>the</strong> Lab Advisor<br />

software. Select Start moni<strong>to</strong>ring.<br />

NOTE<br />

No high voltage can be applied, if <strong>the</strong> front cover is removed. Touching <strong>the</strong> electrodes inside<br />

<strong>the</strong> instrument is safe. However, under <strong>the</strong> following circumstances (usage of <strong>the</strong><br />

alignment center), slow lifter movements with a low force are possible with <strong>the</strong> following<br />

alignment procedure. When you start any lifter movement your hands should be outside <strong>the</strong><br />

instrument.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 213