Using <strong>the</strong> <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> 5<br />

Controlling <strong>the</strong> <strong>Agilent</strong> CE Instrument<br />

Figure 15<br />

Configuration menu<br />

NOTE<br />

Please refer <strong>to</strong> “LAN Configuration” on page 61 for <strong>the</strong> default IP address or manual<br />

configuration of an individual IP address.<br />

After you have double clicked <strong>the</strong> Instrument Online icon, <strong>the</strong> software is loaded<br />

with <strong>the</strong> last method used. The software boots in <strong>the</strong> exact last status.<br />

Do not use <strong>the</strong> instrument or <strong>the</strong> graphical user interface during initialization.<br />

Wait until <strong>the</strong> system has reached <strong>the</strong> ready state.<br />

If <strong>the</strong>re is a vial in tray position 50 (reserved position), <strong>the</strong> system asks you <strong>to</strong><br />

remove that vial before you proceed.<br />

Duplicate vials are put in <strong>the</strong> tray and <strong>the</strong> respective position is brought <strong>to</strong> <strong>the</strong><br />

front. Duplicate vials may occur if vials from <strong>the</strong> tray are loaded in a lifter and<br />

later on ano<strong>the</strong>r vial is loaded in <strong>the</strong> same tray position.<br />

If <strong>the</strong>re are no bottles (electrolyte or waste) in <strong>the</strong> replenishment system or<br />

<strong>the</strong>y are not properly tightened, <strong>the</strong> error LED (red) will come on because no<br />

pressure can be built up. Check <strong>the</strong> logbook for more information.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 83