8 Maintenance<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

9 Push <strong>the</strong> liquid handling module back in<strong>to</strong> <strong>the</strong> instrument as shown in<br />

Figure 70 on page 213. To release it from <strong>the</strong> front position, slightly more<br />

force is needed. Carefully move <strong>the</strong> liquid handling module completely in<strong>to</strong><br />

a position that allows fixation. Please note <strong>the</strong> metal nose from <strong>the</strong> liquid<br />

handling module which has <strong>to</strong> fit in a hoe of <strong>the</strong> back metal sheet as shown<br />

in Figure 69 on page 212.<br />

Metal pin<br />

Connec<strong>to</strong>rs for<br />

water,<br />

pressure,<br />

drainage<br />

Figure 69<br />

Rear side of liquid handling module<br />

NOTE<br />

Take care not <strong>to</strong> damage tubes and flat ribbon cables on <strong>the</strong> right-hand side of <strong>the</strong> liquid<br />

handling module.<br />

Take care that <strong>the</strong> connec<strong>to</strong>rs on <strong>the</strong> left-hand side, for example for <strong>the</strong> drainage, fit in<strong>to</strong><br />

<strong>the</strong>ir cut-out in <strong>the</strong> back.<br />

212 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual