8 Maintenance<br />

Maintenance of <strong>the</strong> Replenishment <strong>System</strong><br />

The Replenishment <strong>System</strong> Will Not Be Used for Some Time<br />

The replenishment system should be flushed first with water and <strong>the</strong>n with<br />

isopropanol.<br />

Please follow <strong>the</strong>se steps.<br />

1 Open <strong>the</strong> Lab Advisor or Instrument Utilities software and start moni<strong>to</strong>ring<br />

<strong>the</strong> respective instrument.<br />

2 In <strong>the</strong> <strong>to</strong>ols section, select <strong>the</strong> CE Control screen and click on Change Bottles <strong>to</strong><br />

release <strong>the</strong> pressure before you open <strong>the</strong> bottles.<br />

3 Take out <strong>the</strong> electrolyte bottle and unscrew <strong>the</strong> cap. The electrolyte bottle<br />

has a frit and tubing inside. The frit must not get dirty.<br />

4 Fill a minimum 300 mL of water in<strong>to</strong> <strong>the</strong> electrolyte bottle. Make sure <strong>the</strong><br />

water is filtered by using a 0.25 µm or 0.45 µm filter depending on its<br />

viscosity.<br />

5 Make sure <strong>the</strong> waste bottle is empty.<br />

6 Cap <strong>the</strong> bottles tightly.<br />

7 Choose Done in <strong>the</strong> Change Bottles dialog box. Wait until <strong>the</strong> pressure has<br />

built up again.<br />

8 To flush <strong>the</strong> internal tubing system from <strong>the</strong> electrolyte bottle <strong>to</strong> <strong>the</strong> waste<br />

bottle, including all valves on this pathway, select Flush Tubes. During this<br />

step <strong>the</strong> replenishment needle itself is not flushed. Set a flushing time of 2<br />

minutes.<br />

9 To clean <strong>the</strong> internal tubing system, including <strong>the</strong> replenishment, select<br />

Clean Tubes. During this step <strong>the</strong> replenishment is used <strong>to</strong> fill and empty a<br />

vial several times. Set <strong>the</strong> number of repetitions <strong>to</strong> 5.<br />

10 When requested <strong>to</strong> do so, place a capped and empty vial in<strong>to</strong> a tray position.<br />

11 Execute <strong>the</strong> clean level sensing operation <strong>to</strong> clean <strong>the</strong> replenishment needle<br />

entirely.<br />

12 Repeat all steps starting with number 2 and use isopropanol instead of<br />

water. Empty <strong>the</strong> waste bottle whenever you fill or refill <strong>the</strong> electrolyte<br />

bottle.<br />

218 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual