5 Using <strong>the</strong> <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong><br />

<strong>Capillary</strong> Electrochroma<strong>to</strong>graphy, CEC<br />

<strong>Capillary</strong> Electrochroma<strong>to</strong>graphy, CEC<br />

<strong>Capillary</strong> electrochroma<strong>to</strong>graphy (CEC) is a fusion of capillary electrophoresis<br />

and capillary chroma<strong>to</strong>graphy. In CEC, solvent is mobilized through a packed<br />

capillary column by electroosmotic flow instead of <strong>the</strong> hydraulic flow used in<br />

liquid chroma<strong>to</strong>graphy LC. Higher efficiencies are achievable because of <strong>the</strong><br />

uniform flow velocity profile and <strong>the</strong> ability <strong>to</strong> use smaller particle sizes in<br />

longer columns than is possible with LC. Operation of CEC requires an equal<br />

application of high pressure <strong>to</strong> <strong>the</strong> inlet and outlet vials in order <strong>to</strong> suppress<br />

bubble formation while avoiding hydrodynamic flow due <strong>to</strong> a pressure<br />

differential.<br />

How <strong>to</strong> Use Your <strong>Agilent</strong> CE Instrument for <strong>Capillary</strong><br />

Electrochroma<strong>to</strong>graphy<br />

Requirements<br />

• External pressure supply (< 15 bar) of oil-free air or nitrogen.<br />

• Push-fit connec<strong>to</strong>r and gas line (supplied in startup kit).<br />

• Outlet lift connection <strong>to</strong> <strong>the</strong> tubing system<br />

NOTE<br />

For external pressure supply, a male adapter fitting <strong>the</strong> instrument’s female adapter and<br />

PTFE tubing (1/8" OD x 1/16" ID PTFE tubing) are part in <strong>the</strong> accessory kit of <strong>the</strong><br />

instrument. Please add a suitable connec<strong>to</strong>r <strong>to</strong> <strong>the</strong> external pressure supply.<br />

The inlet for <strong>the</strong> external high pressure gas is situated at <strong>the</strong> back of <strong>the</strong><br />

instrument.<br />

148 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual