Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

NOTE<br />

Make sure you have disconnected <strong>the</strong> line power cord of <strong>the</strong> <strong>Agilent</strong> CE<br />

instrument before you proceed.<br />

When<br />

Parts required<br />

If arcing or excessive current leakage is observed.<br />

12 mm hexagonal socket screwdriver (included in <strong>the</strong> startup kit)<br />

Screwdriver for slotted screws (included in <strong>the</strong> startup kit)<br />

Screwdriver for cross head screws (included in <strong>the</strong> startup kit)<br />

Wash bottle with isopropanol<br />

Wash bottle with water<br />

Compressed inert gas, oil-free<br />

Prepare <strong>the</strong> <strong>Agilent</strong> CE Instrument<br />

Prior <strong>to</strong> maintenance do <strong>the</strong> following:<br />

1 Make sure no analysis is running or will be started from <strong>the</strong> ChemStation.<br />

2 Start moni<strong>to</strong>ring your CE instrument with <strong>the</strong> Lab Advisor.<br />

3 Select <strong>the</strong> Tools section in <strong>the</strong> instrument panel of <strong>the</strong> Lab Advisor.<br />

4 Select <strong>the</strong> CE control screen and <strong>the</strong>n Maintenance positions.<br />

These steps unload all lifters and move <strong>the</strong>m <strong>to</strong> <strong>the</strong> lowest position. With all<br />

lifters in this position, <strong>the</strong> tray is freely movable. A dialog reports which<br />

lifters were unloaded in case vials were in <strong>the</strong> system.<br />

5 Turn off <strong>the</strong> <strong>Agilent</strong> CE instrument.<br />

6 Disconnect <strong>the</strong> line power cord from <strong>the</strong> <strong>Agilent</strong> CE instrument.<br />

7 Disconnect any of <strong>the</strong> following connection if present: waterbath, external<br />

pressure supply and drainage.<br />

WARNING<br />

Hot lamp<br />

If you were using <strong>the</strong> <strong>Agilent</strong> CE instrument shortly before, <strong>the</strong> lamp may be very hot.<br />

• Avoid <strong>to</strong>uching <strong>the</strong> lamp.<br />

• Wear protective gloves.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 197