4 LAN Configuration<br />

S<strong>to</strong>ring <strong>the</strong> Settings Permanently with Bootp<br />

S<strong>to</strong>ring <strong>the</strong> Settings Permanently with Bootp<br />

If you want <strong>to</strong> change parameters of <strong>the</strong> CE instrument using Bootp follow <strong>the</strong><br />

instructions below.<br />

1 Turn off <strong>the</strong> CE instrument.<br />

2 Change <strong>the</strong> CE instrument's settings for <strong>the</strong> configuration switch <strong>to</strong> “Bootp<br />

& S<strong>to</strong>re” mode, see Table 18 on page 65.<br />

3 Start <strong>the</strong> <strong>Agilent</strong> Bootp Service (see “<strong>Agilent</strong> Bootp Service” on page 75)<br />

and open its window.<br />

4 If required, modify <strong>the</strong> parameters for <strong>the</strong> CE instrument according <strong>to</strong> your<br />

needs using <strong>the</strong> existing configuration.<br />

5 Press OK <strong>to</strong> exit <strong>the</strong> Bootp manager.<br />

6 Now turn on <strong>the</strong> CE instrument and view <strong>the</strong> Bootp server window. After<br />

some time <strong>the</strong> <strong>Agilent</strong> Bootp Service will display <strong>the</strong> request from <strong>the</strong> LAN<br />

interface. The parameters are now s<strong>to</strong>red permanently in <strong>the</strong> non-volatile<br />

memory of <strong>the</strong> CE instrument.<br />

7 Close <strong>the</strong> <strong>Agilent</strong> Bootp Service and turn off <strong>the</strong> CE instrument.<br />

8 Change <strong>the</strong> settings of <strong>the</strong> CE instrument’s configuration switch <strong>to</strong> “Using<br />

S<strong>to</strong>red” mode, see Table 18 on page 65.<br />

9 Power cycle <strong>the</strong> CE instrument. The CE instrument can be accessed now via<br />

LAN without <strong>the</strong> <strong>Agilent</strong> Bootp Service, refer <strong>to</strong> “<strong>Agilent</strong> Bootp Service” on<br />

page 75.<br />

70 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual