Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

NOTE<br />

The screws used for fixation of <strong>the</strong> inlet and outlet pre-puncher are made of plastic. Do not<br />

use metal screws. Metal screws may lead <strong>to</strong> leakage current and arcing which disturbs<br />

applications and may damage <strong>the</strong> instrument.<br />

5 Carefully insert <strong>the</strong> replenishment pre-puncher. Tighten <strong>the</strong> screw <strong>to</strong><br />

secure <strong>the</strong> pre-puncher.<br />

The pre-puncher will assume <strong>the</strong> correct position as <strong>the</strong> screw is tightened.<br />

The replenishment pre-puncher is not connected <strong>to</strong> a tube.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 209