Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

Accessing <strong>the</strong> Liquid Handling Module<br />

CAUTION<br />

Damage <strong>to</strong> tubes and liquid spills<br />

Keeping tubes connected on <strong>the</strong> left side of <strong>the</strong> instrument while pulling out <strong>the</strong> liquid<br />

handling module will damage <strong>the</strong> tubes and might lead <strong>to</strong> liquid spills.<br />

Disconnect any external pressure supply, drainage or connections <strong>to</strong> an external water<br />

bath for tray cooling prior <strong>to</strong> accessing <strong>the</strong> liquid handling module.<br />

CAUTION<br />

Damage <strong>to</strong> <strong>the</strong> instrument<br />

Damage may lead <strong>to</strong> a non-operational instrument.<br />

• Do not lift <strong>the</strong> liquid handling module upwards, outside <strong>the</strong> instrument. The module<br />

is connected <strong>to</strong> <strong>the</strong> main frame by two flat ribbon cables and two flexible tubes.<br />

When lifting it up, <strong>the</strong> liquid handling module might come off <strong>the</strong> guiding rail and <strong>the</strong><br />

connections might be affected.<br />

1 Use a screwdriver <strong>to</strong> unscrew <strong>the</strong> slotted screws holding <strong>the</strong> liquid handling<br />

module in <strong>the</strong> instrument.<br />

2 Take <strong>the</strong> electrolyte and waste bottles out of <strong>the</strong> cabinet. Keep <strong>the</strong>m<br />

connected and place <strong>the</strong>m in front of <strong>the</strong> instrument.<br />

3 Carefully pull <strong>the</strong> liquid handling module out from <strong>the</strong> main frame.<br />

Pull steadily in <strong>the</strong> middle until you reach <strong>the</strong> end position. Carefully avoid<br />

damaging <strong>the</strong> replenishment needle.<br />