Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

Accessing and Removing <strong>the</strong> Pre-punchers<br />

Inlet and outlet lifts are located behind <strong>the</strong> sample tray in <strong>the</strong> rear of <strong>the</strong> liquid<br />

handling module. The replenishment lift is located in <strong>the</strong> front on <strong>the</strong> right<br />

side under <strong>the</strong> replenishment needle.<br />

Outlet lift<br />

Inlet lift<br />

Replenishment needle<br />

Replenishment lift<br />

Figure 62<br />

Inlet and outlet lifts, replenishment lift and replenishment needle<br />

NOTE<br />

Do not <strong>to</strong>uch or bend <strong>the</strong> replenishment needle when handling <strong>the</strong> pre-punchers.<br />

WARNING<br />

Risk of injury due <strong>to</strong> sharp tips<br />

The pre-punchers have very sharp tips.<br />

• Be careful when <strong>to</strong>uching <strong>the</strong> tips of <strong>the</strong> pre-punchers.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 203