Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

5 The insulation plate should be laying flat in position.<br />

The insulation plate is correctly positioned under <strong>the</strong> detec<strong>to</strong>r if <strong>the</strong> <strong>to</strong>p<br />

part of <strong>the</strong> outlet electrode is fits through an oval cut-out.<br />

Figure 68<br />

Alignment of <strong>the</strong> detec<strong>to</strong>r<br />

6 Secure <strong>the</strong> insulation plate by reinstalling <strong>the</strong> two slotted metal screws.<br />

7 Reconnect <strong>the</strong> blue ground wire and fasten <strong>the</strong> metal screw.<br />

8 Insert <strong>the</strong> inlet electrode in<strong>to</strong> <strong>the</strong> ring connec<strong>to</strong>r of <strong>the</strong> red high voltage<br />

wire. Thread <strong>the</strong> inlet electrode (and <strong>the</strong> ring connec<strong>to</strong>r) in<strong>to</strong> <strong>the</strong> insulation<br />

plate by hand, <strong>the</strong>n tighten <strong>the</strong> inlet electrode with <strong>the</strong> 12 mm hexagonal<br />

socket screwdriver.<br />

Do not overtighten.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 211