2 Site Requirements and Specifications<br />

Site Requirements<br />

Site Requirements<br />

A suitable environment is important <strong>to</strong> ensure optimal performance of <strong>the</strong><br />

<strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> instrument.<br />

Before you begin installation, check carefully that <strong>the</strong> place you have chosen<br />

meets <strong>the</strong> requirements below.<br />

Power Consideration<br />

The instrument power supply has wide ranging capabilities and accepts any<br />

standard line voltage in <strong>the</strong> range mentioned in Table 2 on page 21.<br />

Consequently, <strong>the</strong>re is no voltage selec<strong>to</strong>r on <strong>the</strong> back of <strong>the</strong> detec<strong>to</strong>r.<br />

WARNING<br />

The module is partially energized when switched off, as long as <strong>the</strong> power cord is<br />

plugged in.<br />

Repair or maintenance work on <strong>the</strong> module can lead <strong>to</strong> personal injuries, for<br />

example a shock hazard, when <strong>the</strong> cover is opened and <strong>the</strong> module is connected <strong>to</strong><br />

<strong>the</strong> power.<br />

• Make sure that it is always possible <strong>to</strong> access <strong>the</strong> power plug.<br />

• Remove <strong>the</strong> power cable from <strong>the</strong> instrument before opening <strong>the</strong> cover.<br />

• Do not connect <strong>the</strong> power cable <strong>to</strong> <strong>the</strong> instrument while <strong>the</strong> electronic box in <strong>the</strong><br />

back of <strong>the</strong> instrument is opened or side panels are removed.<br />

WARNING<br />

Incorrect line voltage on <strong>the</strong> instrument<br />

Shock hazard or damage <strong>to</strong> your instrument can result if <strong>the</strong> device is connected <strong>to</strong> a<br />

line voltage that is higher than specified.<br />

• Connect your instrument <strong>to</strong> <strong>the</strong> specified line voltage.<br />