Maintenance 8<br />

Early Maintenance Feedback (EMF)<br />

Setting <strong>the</strong> EMF Limits<br />

The setting for <strong>the</strong> EMF limits must be optimized over one or two maintenance<br />

cycles. Initially, no EMF limit should be set. When instrument performance<br />

indicates maintenance is necessary, take note of <strong>the</strong> values displayed by <strong>the</strong><br />

instrument counters. Enter <strong>the</strong>se values (or values slightly less than <strong>the</strong><br />

displayed values) as EMF limits and <strong>the</strong>n reset <strong>the</strong> EMF counters <strong>to</strong> zero. The<br />

next time <strong>the</strong> EMF counters exceed <strong>the</strong> new EMF limits, <strong>the</strong> EMF flag will be<br />

displayed, providing a reminder that maintenance needs <strong>to</strong> be scheduled.<br />

NOTE<br />

Please introduce self-made EMF counters <strong>to</strong> remind you <strong>to</strong> frequently clean <strong>the</strong> electrodes,<br />

pre-punchers and insulation plate.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 195