Using <strong>the</strong> <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> 5<br />

<strong>Capillary</strong> <strong>Electrophoresis</strong>, Plus High Pressure CE+p<br />

Timetable controlled changes in <strong>the</strong> pressure settings are:<br />

• OFF: Pressure usage is switched off.<br />

• Internal, pressure (-100 mbar <strong>to</strong> +100 mbar): Can be applied <strong>to</strong> <strong>the</strong> inlet<br />

vial.<br />

• Flush: Standard flush procedure is applied with an internally generated<br />

pressure of approximately 950 mbar.<br />

• External: Allows pressurizing <strong>the</strong> inlet vial and/or <strong>the</strong> outlet vial. External<br />

pressure is reduced <strong>to</strong> <strong>the</strong> set value between 2 and 12 bar and applied in a<br />

controlled manner.<br />

• Raw External Pressure: Allows pressurizing <strong>the</strong> inlet vial, and/or <strong>the</strong> outlet<br />

vial. The external pressure up <strong>to</strong> 15 bar can be applied directly without<br />

reduction.<br />

CE+p Control Using <strong>the</strong> CE Diagram<br />

High pressure may be applied independently of <strong>the</strong> method for capillary<br />

conditioning or cleaning by clicking <strong>the</strong> external pressure icon in <strong>the</strong> CE<br />

diagram (see Figure 39 on page 142). This will cause a parameter setting field<br />

<strong>to</strong> appear (see Figure 41 on page 145). From this field you may chose <strong>to</strong> apply<br />

pressure <strong>to</strong> <strong>the</strong> inlet, outlet or both vials, and also whe<strong>the</strong>r <strong>to</strong> simultaneously<br />

apply a voltage of -30 <strong>to</strong> 30 kV for a given time.<br />