8 Maintenance<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

Cleaning <strong>the</strong> Pre-punchers<br />

• Check for damaged or bent tips before cleaning <strong>the</strong> pre-punchers.<br />

If <strong>the</strong> tips are damaged, replace <strong>the</strong> pre-punchers.<br />

• Check for grooves in <strong>the</strong> funnels of <strong>the</strong> pre-punchers.<br />

Grooves are a sign of improper alignment of <strong>the</strong> electrodes and may lead <strong>to</strong><br />

improper sealing. Improper sealing may lead <strong>to</strong> problems during flushing,<br />

injection or CEC applications. Use <strong>the</strong> alignment center of <strong>the</strong> Lab Advisor<br />

<strong>to</strong> align and check <strong>the</strong> installation. If <strong>the</strong> funnel shows grooves, replace <strong>the</strong><br />

pre-punchers.<br />

WARNING<br />

Risk of injury due <strong>to</strong> sharp tips<br />

The pre-punchers have very sharp tips.<br />

• Be careful when <strong>to</strong>uching <strong>the</strong> tips of <strong>the</strong> pre-punchers.<br />

1 Flush <strong>the</strong> pre-punchers with water. Watch for salt crystals. All salt crystals<br />

must be removed. Take particular care with <strong>the</strong> teflon tube connec<strong>to</strong>r.<br />

Figure 65<br />

Flushing pre-punchers<br />

2 Flush <strong>the</strong> pre-punchers with water and isopropanol.<br />

3 Use compressed air <strong>to</strong> dry <strong>the</strong> pre-punchers and remove all traces of liquid.<br />

Do not forget <strong>to</strong> dry <strong>the</strong> connected tubes as well.<br />