5 Using <strong>the</strong> <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong><br />

<strong>Capillary</strong> <strong>Electrophoresis</strong>, CE<br />

Figure 21<br />

DAD Method Setup<br />

Selecting OK transfers <strong>the</strong> parameters <strong>to</strong> <strong>the</strong> instruments and closes <strong>the</strong> dialog.<br />

Choosing Apply only transfers <strong>the</strong> current parameter <strong>to</strong> <strong>the</strong> instrument. Save<br />

<strong>the</strong> edited method, if appropriate.<br />

If you do not save <strong>the</strong> method <strong>to</strong> <strong>the</strong> hard disk, this changed method is<br />

executed when choosing Run Method. When you refer <strong>to</strong> a method in a<br />

sequence, <strong>the</strong> method is always loaded from <strong>the</strong> disk. So if you did not save<br />

your changes using Save Method, <strong>the</strong> original method from <strong>the</strong> hard disk is<br />

used.<br />

Method parameters are also available if Method/Edit Entire Method... is chosen.<br />

This dialog guides you through fur<strong>the</strong>r parameters associated with a method<br />

such as<br />

94 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual