8 Maintenance<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

1 Remove <strong>the</strong> plastic screw which secures <strong>the</strong> pre-punchers using a Pozidriv<br />

screwdriver.<br />

Teflon tubing<br />

Inlet pre-puncher<br />

Screw, plastic<br />

Figure 63<br />

Plastic screw of <strong>the</strong> pre-puncher<br />

2 The pre-puncher has a hole which can be used <strong>to</strong> gently lever it out using a<br />

paper clip. Carefully pull <strong>the</strong> pre-puncher out of <strong>the</strong> inlet, outlet and<br />

replenishment lift.<br />

3 Unscrew <strong>the</strong> teflon inlet pre-puncher tubing (labeled 7 on <strong>the</strong> tubing), <strong>the</strong><br />

outlet pre-puncher tubing (labeled 8). The pre-puncher for <strong>the</strong><br />

replenishment lift is not connected <strong>to</strong> a tube.<br />

NOTE<br />

The outlet lifter tubing (labeled 8) might be physically disconnected from <strong>the</strong> pressure<br />

system. This status is advantageous in case no high pressure applications are used. This<br />

connection is required for applications in CE+p and CEC mode.<br />

NOTE<br />

The teflon tube does NOT need <strong>to</strong> be removed from <strong>the</strong> pre-punchers for cleaning. If <strong>the</strong><br />

connection between <strong>the</strong> tube and <strong>the</strong> pre-puncher is not tight, since <strong>the</strong> tube became<br />

wider, shorten <strong>the</strong> tube and shift it on <strong>the</strong> pre-puncher again.<br />