8 Maintenance<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

Reinstalling <strong>the</strong> Electrodes and <strong>the</strong> Insulation Plate<br />

1 Reinstall one electrode in <strong>the</strong> outlet electrode position on <strong>the</strong> insulation<br />

plate.<br />

2 Thread <strong>the</strong> electrode in<strong>to</strong> <strong>the</strong> outlet position on <strong>the</strong> insulation plate by hand<br />

first.<br />

3 Tighten <strong>the</strong> electrode with <strong>the</strong> 12 mm hexagonal socket screwdriver.<br />

Do not overtighten. Do not bend <strong>the</strong> electrodes.<br />

4 Holding <strong>the</strong> insulation plate by <strong>the</strong> cassette holder, slide <strong>the</strong> insulation<br />

plate under <strong>the</strong> detec<strong>to</strong>r (1.), <strong>the</strong>n tilt it up in<strong>to</strong> place (2.) and push it<br />

completely <strong>to</strong> <strong>the</strong> left (3.).<br />

3. Left shift<br />

2. Rotation<br />

1. Insert<br />

Figure 67<br />

Installation of <strong>the</strong> insulation plate<br />