8 Maintenance<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

13 Open <strong>the</strong> CE Control screen under Tools in <strong>the</strong> Lab Advisor software. Select<br />

<strong>the</strong> Alignment Center. The alignment procedure can be started with <strong>the</strong> front<br />

cover un-installed.<br />

14 Follow <strong>the</strong> procedure <strong>to</strong> align <strong>the</strong> inlet and outlet electrode and <strong>the</strong><br />

replenishment needle.<br />

The procedure consists of a prepare step and an alignment check.<br />

The alignment is done while <strong>the</strong> pre-puncher of <strong>the</strong> respective lift is raised<br />

until it is near <strong>the</strong> electrode or needle tip (prepare-but<strong>to</strong>n). Now you can<br />

insert <strong>the</strong> capillary of <strong>the</strong> electrode alignment <strong>to</strong>ol from <strong>the</strong> funnel of <strong>the</strong><br />

electrode. With careful puncture you can check if <strong>the</strong> electrode exit points<br />

fit exactly <strong>to</strong> <strong>the</strong> pre-puncher funnel exit. The electrode or needle might<br />

need careful manual bending <strong>to</strong> make it point <strong>to</strong>wards <strong>the</strong> exect center of<br />

<strong>the</strong> pre-puncher funnel.<br />

During <strong>the</strong> alignment check (check-but<strong>to</strong>n), <strong>the</strong> electrode or replenishment<br />

needle passes through <strong>the</strong> funnel several times. An improperly aligned<br />

electrode slides along <strong>the</strong> pre-puncher funnel. The orientation is optimum if<br />

<strong>the</strong> electrode does not move sidewards during this dynamic check.<br />

NOTE<br />

Although electrodes will most likely not need alignment thanks <strong>to</strong> <strong>the</strong>ir design, you should<br />

still perform <strong>the</strong> check alignment procedure.<br />

NOTE<br />

The electrodes are fragile. Manual bending during alignment should be done only with<br />

great care. Electrodes must be kept straight. Avoid curving.<br />