5 Using <strong>the</strong> <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong><br />

<strong>Capillary</strong> Electrochroma<strong>to</strong>graphy, CEC<br />

Preparing <strong>the</strong> <strong>Agilent</strong> CE Instrument for High Pressure Use<br />

1 Ensure that <strong>the</strong> gas source is switched off prior <strong>to</strong> connection.<br />

2 Connect <strong>the</strong> push-fit connec<strong>to</strong>r and gas line <strong>to</strong> <strong>the</strong> recep<strong>to</strong>r on <strong>the</strong> rear of<br />

<strong>the</strong> <strong>Agilent</strong> CE instrument.<br />

3 Connect <strong>the</strong> gas line <strong>to</strong> <strong>the</strong> source.<br />

4 Switch on <strong>the</strong> gas supply.<br />

5 Select under Instrument in <strong>the</strong> CE diagram <strong>the</strong> Instrument Configuration Menu,<br />

see Figure 44.<br />

Figure 44<br />

Configuration menu<br />

150 <strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual