8 Maintenance<br />

Early Maintenance Feedback (EMF)<br />

Early Maintenance Feedback (EMF)<br />

Maintenance requires <strong>the</strong> exchange of components which are subject <strong>to</strong> wear<br />

or stress. Ideally, <strong>the</strong> frequency at which components are exchanged or<br />

cleaned should be based on <strong>the</strong> intensity of usage of <strong>the</strong> instrument and <strong>the</strong><br />

analytical conditions, and not on a predefined time interval. The Early<br />

Maintenance Feedback (EMF) feature moni<strong>to</strong>rs <strong>the</strong> usage of specific<br />

components in <strong>the</strong> instrument, and provides feedback when <strong>the</strong><br />

user-selectable limits have been exceeded. The visual feedback in <strong>the</strong> user<br />

interface provides an indication that maintenance procedures should be<br />

scheduled.<br />

EMF Counters<br />

The CE instrument provides EMF counters for <strong>the</strong> CE part and <strong>the</strong> detec<strong>to</strong>r.<br />

The counters increment with instrument use, and can be assigned a maximum<br />

limit which provides visual feedback in <strong>the</strong> user interface when <strong>the</strong> limit is<br />

exceeded. The counters can be reset <strong>to</strong> zero after <strong>the</strong> related parts are<br />

exchanged. The CE instrument provides <strong>the</strong> following EMF counters:<br />

• Analysis start counter<br />

• Accumulated air pump on time<br />

• Accumulated UV lamp on time<br />

• Number of ignitions<br />

Using <strong>the</strong> EMF Counters<br />

The user-settable EMF limits for <strong>the</strong> EMF counters enable <strong>the</strong> Early<br />

Maintenance Feedback <strong>to</strong> be adapted <strong>to</strong> specific user requirements. The useful<br />

lamp burn time is dependent on <strong>the</strong> requirements for <strong>the</strong> analysis (high or low<br />

sensitivity analysis, wavelength, etc.), <strong>the</strong>refore, <strong>the</strong> definition of <strong>the</strong><br />

maximum limits need <strong>to</strong> be determined based on <strong>the</strong> specific operating<br />

conditions of <strong>the</strong> instrument.<br />