Troubleshooting and Diagnostics 6<br />

CE Problem Solving<br />

Empty Injection Vial<br />

Access <strong>the</strong> vial position specified in <strong>the</strong> Sample info or <strong>the</strong> Sequence table using<br />

<strong>the</strong> Get vial function in <strong>the</strong> Tray control screen.<br />

• Check if <strong>the</strong> vial is filled.<br />

Dirt in <strong>the</strong> Vials<br />

Clean all vials prior <strong>to</strong> use. First, flush <strong>the</strong>m with water followed by, for<br />

example, some isopropanol.<br />

• Use compressed air <strong>to</strong> dry <strong>the</strong> vials inside.<br />

Air Bubble in Injection Vial or Run Buffer Vials<br />

When filling sample or buffers in<strong>to</strong> <strong>the</strong> vial, an air bubble is sometimes<br />

formed. This happens very frequently when using microvials. Air bubbles can<br />

cause air <strong>to</strong> be injected instead of <strong>the</strong> sample.<br />

• To avoid air bubbles, carefully fill vials starting from <strong>the</strong> bot<strong>to</strong>m.<br />

• If you cannot avoid <strong>the</strong> formation of air bubbles, adjust <strong>the</strong> lift offset in <strong>the</strong><br />

Home values dialog box so that <strong>the</strong> capillary doesn't reach in<strong>to</strong> <strong>the</strong> bubbles.<br />

Air Bubble in <strong>Capillary</strong><br />

Small air bubbles in <strong>the</strong> capillary may cause spikes in your electropherogram.<br />

Larger air bubbles expand so much during analysis that <strong>the</strong> current drops<br />

below <strong>the</strong> lower alarm limit.<br />

• Flush your capillary with buffer <strong>to</strong> remove air bubbles.<br />

<strong>Agilent</strong> <strong>7100</strong> <strong>Capillary</strong> <strong>Electrophoresis</strong> <strong>System</strong> User Manual 167