Maintenance 8<br />

Cleaning <strong>the</strong> Electrodes, Pre-punchers and Insulation Plate<br />

NOTE<br />

If <strong>the</strong>re is an accumulation of dirt in <strong>the</strong> pre-puncher funnel, put <strong>the</strong> pre-punchers and<br />

connecting tubing in a beaker of water and treat it in an ultrasonic bath for five minutes<br />

followed by steps 2 and 3 in this section.<br />

NOTE<br />

Residual liquid in connecting tubes may be transported <strong>to</strong> <strong>the</strong> air valve blocks and result in<br />

damage. Such liquid plugs may also negatively influence injection reproducibility.<br />

Cleaning <strong>the</strong> Insulation Plate<br />

Salt crystals, films of substances such as buffer or gel that allow moisture will<br />

lead <strong>to</strong> increased leakage current or even arcing. The insulation plate must be<br />

free of such contamination.<br />

1 Moisten a tissue with water and use it <strong>to</strong> clean <strong>the</strong> insulation plate.<br />

2 Moisten a tissue with isopropanol and use it <strong>to</strong> clean <strong>the</strong> insulation plate.<br />

3 Dry <strong>the</strong> insulation plate with a tissue and compressed air.<br />