Analyzing the Study• Manual Ct – Instructs the software to use the baseline <strong>and</strong> threshold valuesspecified in the Analysis Settings dialog box to calculate C T s for the selecteddetectors.The software provides two methods for manually setting the baseline <strong>and</strong>threshold values for the detector:– If you know the values that you want to use, enter them into the appropriatefields of the Analysis Settings dialog box.– If you want to set the baseline <strong>and</strong> threshold using the controls of theAmplification Plot tab, leave the fields empty <strong>and</strong> set the values later asdescribed on page 6-46.Replicate <strong>and</strong>Outlier Removalfor Study SettingsThe settings of this group box determine how the software identifies <strong>and</strong> removesoutlier data for all detectors in the study.Automatic OutlierRemoval check boxOmit replicates whose ∆C T