Chapter 7Maintaining the InstrumentPreparing theVerification Plateor TGF-β CardPreparing a TaqMan RNase P Instrument Verification Plate1. Remove the TaqMan RNase P Instrument Verification Plate from the freezer<strong>and</strong> allow it to thaw to room temperature.2. Briefly centrifuge the TaqMan RNase P Instrument Verification Plate.3. Continue with “Preparing the Plate Document” on page 7-33.Preparing a TGF-β Card4. Remove the TGF-β card from the refrigerator <strong>and</strong> allow it to warm to roomtemperature.5. In a sterile 2-mL tube, prepare the following <strong>PCR</strong> reaction mixture:ReagentVolume (µL)TaqMan 2✕ Universal <strong>PCR</strong> Master Mix 500Human cDNA 50RNase/DNase-free water 450Total Volume 10006. Tap the tube gently to mix.7. Pipette 100 µL of the reaction mix into each of the eight fill reservoirs.8. Continue with “Preparing the Plate Document” on page 7-33.7-32 <strong>Applied</strong> <strong>Biosystems</strong> <strong>7900HT</strong> <strong>Fast</strong> <strong>Real</strong>-<strong>Time</strong> <strong>PCR</strong> <strong>System</strong> <strong>and</strong> <strong>SDS</strong> Enterprise Database User GuideDRAFTSeptember 1, 2004 11:39 am, CH_Maintenance.fm