Software <strong>and</strong> <strong>7900HT</strong> InstrumentTable 8-4Troubleshooting Software <strong>and</strong> Computer ProblemsObservation Possible Cause Recommended ActionCommunication errorThermal cycler errorsAutomation ControllerSoftware cannot find a platedocument fileDialog box does not respondto mouse clicks or keystrokesCables are connectedincorrectlySample block modulenot fully engagedFile not in correctlocationJava Runtime ErrorCheck cable connections <strong>and</strong>COM port setup. See“Instrument Connections” onpage 1-10.Reseat the sample blockmodule as explained“Replacing the Sample Block”on page 7-6.Remove file entry from platequeue <strong>and</strong> add the file to theplate queue again.Click the close box of the dialogbox to close it.Run will not start No calibration file Perform background <strong>and</strong> puredye runs.No background data inSee “Performing a Backgroundcalibration fileRun” on page 7-16 <strong>and</strong>(background run has“Performing a Pure Dye Run”not been performed)on page 7-20.No pure dye data incalibration file(pure dye run has notbeen performed)Calibration file doesnot contain pure dyedata for a dye used onthe plate documentCalibration file wascreated on anotherinstrumentDisk drive containingthe plate documenthas less than 50 MB offree spaceHeated cover cannotreach runningtemperature becauseno plate loadedCheck the capacity of thedestination drive. If less than50 MB of free space remains,remove or archive existing datafiles (see page 7-54).Open the instrument tray <strong>and</strong>check that the instrumentcontains a plate.Instrument traycontains a plateOutput stack containsa plate or platesRemove all plates from theoutput stack of the PlateH<strong>and</strong>ler before starting thequeue.<strong>Applied</strong> <strong>Biosystems</strong> <strong>7900HT</strong> <strong>Fast</strong> <strong>Real</strong>-<strong>Time</strong> <strong>PCR</strong> <strong>System</strong> <strong>and</strong> <strong>SDS</strong> Enterprise Database User Guide 8-15DRAFTSeptember 1, 2004 11:39 am, CH_Trouble.fm

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!