Replacing the Sample Block9. Loosen the thumbscrew securing the sample block locking bar to the instrumentchassis (may be a 5/32-inch hex bolt on some instruments).GR2025bGR2024bthumbscrew10. Lift the sample block locking bar up <strong>and</strong> out of the instrument.11. Remove the sample block from the instrument:a. Rotate the release lever at the base of the sample block 90 degrees.b. Being careful not to damage the heat sinks on the bottom of the sampleblock, slide the sample block out of the instrument <strong>and</strong> place it on a clean,level surface.GR2027GR2025b<strong>Applied</strong> <strong>Biosystems</strong> <strong>7900HT</strong> <strong>Fast</strong> <strong>Real</strong>-<strong>Time</strong> <strong>PCR</strong> <strong>System</strong> <strong>and</strong> <strong>SDS</strong> Enterprise Database User Guide 7-9DRAFTSeptember 1, 2004 11:39 am, CH_Maintenance.fm