Chapter 7Maintaining the InstrumentReplacing the Sample BlockWhen to PerformSample BlockInstallationWorkflowIMPORTANT! Before changing the sample block, perform all required upgrades tothe <strong>SDS</strong> software <strong>and</strong> instrument firmware. Failure to update the software can makethe instrument inoperable or result in damage to instrument components.You need to remove the <strong>7900HT</strong> instrument sample block when you:• Decontaminate the wells of the sample block (see page 7-14)• Change sample block formatsIMPORTANT! Always run a background plate after installing the sample block.Unless instructed to do otherwise, adhere to the following guidelines whenexchanging sample block modules of different formats.To install a sample block module:1. Perform all required software <strong>and</strong> firmware upgrades to the <strong>Applied</strong> <strong>Biosystems</strong><strong>7900HT</strong> <strong>Fast</strong> <strong>Real</strong>-<strong>Time</strong> <strong>PCR</strong> <strong>System</strong> (see page 7-55).2. Remove the existing sample block (see page 7-7).3. Install the new sample block (see page 7-10).4. Change the plate adapter (see page 7-12).5. Run a background plate to check the sample block for contamination(see page 7-16).If you are changing block formats or installing a new block, also do the following:IMPORTANT! Perform pure dye <strong>and</strong> background runs after replacing the sampleblock, even if you are replacing a sample block module with another block of thesame format.6. Run a pure dye plate or microfluidic pure dye card to create the pure spectracalibration values for the new format (see page 7-20).7. Run the appropriate TaqMan ® RNase P Instrument Verification Plate or aTaqMan ® Low Density <strong>7900HT</strong> Installation Array (TGF-β card) to confirm theproper operation of the sample block (see page 7-30).If you are using an Automation Accessory, also do the following:8. If changing consumable formats (for example, when replacing a St<strong>and</strong>ard384-Well Block with a St<strong>and</strong>ard 96-Well Block, a <strong>Fast</strong> 96-Well Block, or a<strong>7900HT</strong> <strong>System</strong> TaqMan ® Low Density Array Upgrade), adjust the plate sensorswitch on the Plate H<strong>and</strong>ler arm for the new plates (see page 7-37).9. Align the Zymark ® Twister Microplate H<strong>and</strong>ler fixed-position bar code readerfor the new plate format (see page 7-41).IMPORTANT! You must align the Plate H<strong>and</strong>ler to only the Instrument position(Zymark position 2) as explained on pages 7-41 to 7-44.10. Align the fixed-position bar code reader for the new plate format(see page 7-49).7-6 <strong>Applied</strong> <strong>Biosystems</strong> <strong>7900HT</strong> <strong>Fast</strong> <strong>Real</strong>-<strong>Time</strong> <strong>PCR</strong> <strong>System</strong> <strong>and</strong> <strong>SDS</strong> Enterprise Database User GuideDRAFTSeptember 1, 2004 11:39 am, CH_Maintenance.fm

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!