Aligning the Plate H<strong>and</strong>ler5. Using the Rotary Adjustment arrows, adjust the rotational position of thegripper so that it is centered over the input stack <strong>and</strong> will not contact the sideswhen lowered.• To move the Plate H<strong>and</strong>ler arm clockwise, click the up arrow.• To move the Plate H<strong>and</strong>ler arm counter-clockwise, click the down arrow.Up arrow (moves the arm clockwise)Down arrow (moves the arm counter-clockwise)6. Using the Vertical Positioning comm<strong>and</strong>s, carefully lower the Plate H<strong>and</strong>lerarm into the stack. Adjust the Rotary Adjustment value as needed to center thegripper inside the stack.7. After the gripper is centered inside the stack, click .The Plate H<strong>and</strong>ler arm lowers upon the plate.Confirm the following:• the plate is in the middle of the gripper span• the plate sensor switch is contacting the plate• the gripper <strong>and</strong> plate do not contact the side of the stack8. Click .The gripper grips the plate between its fingers.9. Select .The Plate H<strong>and</strong>ler raises the arm to its highest position. If the plate contacts thesides of the stack, re-adjust the rotary position of the Plate H<strong>and</strong>ler arm until theplate moves freely in the stack.Note: Contact between the plate <strong>and</strong> the stack or all stacks may be unavoidable.However, try to minimize the contact as much as possible.10. Using the Vertical Positioning comm<strong>and</strong>s, raise <strong>and</strong> lower the Plate H<strong>and</strong>ler armseveral times to check the alignment.11. Lower the Plate H<strong>and</strong>ler arm to the bottom of the plate stack, click ,<strong>and</strong> click .The software records the rotary position for the Zymark position 4(input stack 1).12. Click .The gripper releases the plate.<strong>Applied</strong> <strong>Biosystems</strong> <strong>7900HT</strong> <strong>Fast</strong> <strong>Real</strong>-<strong>Time</strong> <strong>PCR</strong> <strong>System</strong> <strong>and</strong> <strong>SDS</strong> Enterprise Database User Guide 7-43DRAFTSeptember 1, 2004 11:39 am, CH_Maintenance.fm

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!