Chapter 7Maintaining the Instrument10. Exit the Zymark Twister Software.a. Click .b. Click Quit Application to close the software.A defect in the Zymark Twister Software can cause portions of the program to persistin memory even after the software has been closed. Because the Zymark TwisterSoftware conflicts with the <strong>SDS</strong> software, the residual elements of the software mustbe closed inside the Windows Task Manager before continuing.1. Confirm that the stack has closed by viewing the Task Manager.a. Press the Crtl + Alt + Del keys in unison.b. In the Windows Security dialog box, click Task Manager.c. In the Task Manager dialog box, confirm that the software has closed bylooking for the Zymark Twister Software entry in the Task list. If thesoftware is still running, click the software entry <strong>and</strong> click Close to exit theremaining software.d. Select File > Exit.7-40 <strong>Applied</strong> <strong>Biosystems</strong> <strong>7900HT</strong> <strong>Fast</strong> <strong>Real</strong>-<strong>Time</strong> <strong>PCR</strong> <strong>System</strong> <strong>and</strong> <strong>SDS</strong> Enterprise Database User GuideDRAFTSeptember 1, 2004 11:39 am, CH_Maintenance.fm