<strong>Tsunami</strong><br />
Mode Locking the <strong>Laser</strong><br />
6. When the laser is optimized in the vertical direction, repeat steps 3<br />
through 5 using the horizontal controls.<br />
7. If necessary, recenter the beam on Mlo. Carefully adjust M3 (and M4<br />
and M5 if necessary), then adjust Mlo to maximize output power.<br />
8. To recenter the beam on MI, carefully adjust MP. Then again adjust MI<br />
to maximize output power.<br />
9. Repeat this entire procedure, starting at step 3, until no further<br />
increase in power is noted.<br />
This completes the procedure for "walking the beam" to optimize laser<br />
output. Continue on to "Mode Locking the <strong>Laser</strong>."<br />
Proceed to mode lock the laser once the pump laser and the Model 3955<br />
are wanned up.<br />
1. Push the STATUS mode locker enable button (the ENABLE LED should<br />
be lit).<br />
The two TEMPERATURE indicators on the <strong>Tsunami</strong> laser control panel<br />
turn from red to green when the AOM and GTI (ps configuration)<br />
heaters stabilize. Proceed to step 2 when the indicators are green.<br />
AJZashing red indicator means either: (a) the GTI is not installed<br />
(normal for fs configuration) or (b) there is an unplugged cable.<br />
2. To measure the <strong>Tsunami</strong> output pulse width, use an autocorrelator<br />
such as a Spectra-Physics <strong>Laser</strong>s Model 409. Avoid saturating the<br />
autocorrelator (refer to the autocorrelator manual).<br />
For a brief discussion on autocorrelation and measuring pulse widths,<br />
refer to appendix B, "Pulse Width Measurement."<br />
Skip to step 4 if you just optimized output power in the previous section.<br />
3. Maximize <strong>Tsunami</strong> output power.<br />
Place a power meter in the <strong>Tsunami</strong> beam path and adjust Mlo and MI<br />
for maximum output power.<br />
4. Adjust the autocorrelator for maximum signal.<br />
5. Adjust the mode locker fine PHASE control for shortest pulse.<br />
If laser output pulses are not present, adjust the fine PHASE control on<br />
the Model 3955 electronics module until either a short pulse appears<br />
on the oscilloscope or the autocorrelator signal is maximized.<br />
If you are using cables other than those shipped with the system, or if<br />
you converted the laser from a ps to a fs configuration, or vice versa,<br />
or if you changed optics sets, the phase signal may have moved<br />
outside the range of the fine PHASE control. If this seems to be the