<strong>Tsunami</strong><br />

Mode Locking the <strong>Laser</strong><br />

6. When the laser is optimized in the vertical direction, repeat steps 3<br />

through 5 using the horizontal controls.<br />

7. If necessary, recenter the beam on Mlo. Carefully adjust M3 (and M4<br />

and M5 if necessary), then adjust Mlo to maximize output power.<br />

8. To recenter the beam on MI, carefully adjust MP. Then again adjust MI<br />

to maximize output power.<br />

9. Repeat this entire procedure, starting at step 3, until no further<br />

increase in power is noted.<br />

This completes the procedure for "walking the beam" to optimize laser<br />

output. Continue on to "Mode Locking the <strong>Laser</strong>."<br />

Proceed to mode lock the laser once the pump laser and the Model 3955<br />

are wanned up.<br />

1. Push the STATUS mode locker enable button (the ENABLE LED should<br />

be lit).<br />

The two TEMPERATURE indicators on the <strong>Tsunami</strong> laser control panel<br />

turn from red to green when the AOM and GTI (ps configuration)<br />

heaters stabilize. Proceed to step 2 when the indicators are green.<br />

AJZashing red indicator means either: (a) the GTI is not installed<br />

(normal for fs configuration) or (b) there is an unplugged cable.<br />

2. To measure the <strong>Tsunami</strong> output pulse width, use an autocorrelator<br />

such as a Spectra-Physics <strong>Laser</strong>s Model 409. Avoid saturating the<br />

autocorrelator (refer to the autocorrelator manual).<br />

For a brief discussion on autocorrelation and measuring pulse widths,<br />

refer to appendix B, "Pulse Width Measurement."<br />

Skip to step 4 if you just optimized output power in the previous section.<br />

3. Maximize <strong>Tsunami</strong> output power.<br />

Place a power meter in the <strong>Tsunami</strong> beam path and adjust Mlo and MI<br />

for maximum output power.<br />

4. Adjust the autocorrelator for maximum signal.<br />

5. Adjust the mode locker fine PHASE control for shortest pulse.<br />

If laser output pulses are not present, adjust the fine PHASE control on<br />

the Model 3955 electronics module until either a short pulse appears<br />

on the oscilloscope or the autocorrelator signal is maximized.<br />

If you are using cables other than those shipped with the system, or if<br />

you converted the laser from a ps to a fs configuration, or vice versa,<br />

or if you changed optics sets, the phase signal may have moved<br />

outside the range of the fine PHASE control. If this seems to be the