<strong>Tsunami</strong><br />

8. Open the pump laser shutter, and adjust Mlo to restore maximum power.<br />

9. Close the pump laser shutter. -<br />

10. Remove and clean the M3 focus mirror as you did Mlo, cleaning the<br />

pump beam input surface first, then the intracavity surface.<br />

11. Install the mirror holder into its mount.<br />

12. Replace the beam shield removed for cleaning PI.<br />

13. Open the pump laser shutter and adjust M3 vertically and horizontally<br />

for maximum output power. Do not adjust the slide.<br />

High Reflector MI, Focus Mirror M2,<br />

Fold Mirrors M4 and M5, and Prism Mirrors ME - Mg<br />

1. Close the pump laser shutter.<br />

2. Remove each mirror, one at a time, starting at M1 and ending at M9<br />

(but not M3), but do not remove the element from its holder. Refer to<br />

Figure 4-3 and Figure 4-4<br />

3. Wearing clean gloves or finger cots, clean each mirror surface using<br />

the "drop and drag" method outlined in step 2 of "General Procedures<br />

for Cleaning Optics."<br />

4. After each mirror is cleaned, screw the mirror holder back in place.<br />

Take care not to cross-thread the holder.<br />

5. Open the pump laser shutter and adjust the optic just cleaned for maximum<br />

output power.<br />

6. Close the pump laser shutter again before removing the next mirror.<br />

Continue in this manner until either specified power is reached, or all the<br />

optics are clean.<br />

GTI (ps configuration)<br />

1. Close the pump laser shutter.<br />

2. Remove the GTI for cleaning. Clean the exposed optic using a folded<br />

tissue and hemostat as described in step 3 of "General Procedure for<br />

Cleaning Optics."<br />

WA RNlNG<br />

Do not disassemble the GTI for cleaning. You will damage it and void<br />

the warranty.<br />

3. Open the pump laser shutter and adjust the GTI vertically and horizontally<br />

for maximum output power.