Tsunami - Beckman Institute Laser Resource Center
Tsunami - Beckman Institute Laser Resource Center
Tsunami - Beckman Institute Laser Resource Center
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
4<br />
<strong>Tsunami</strong><br />
Beam %<br />
2 mm (0.08 in.)<br />
Prism<br />
Figure 7-4: Placement of Pr, and Pr4 in beam<br />
5. Move prisms Prl and Pr4 SO that the beam passes through them about<br />
2 mm from their apices (Figure 7-4). For x-long wavelengths, ensure<br />
the beam is not clipped by these prisms.<br />
6. Use the prism dispersion compensation control to adjust Pr2 and Pr3<br />
so the beam also passes through them about 2 mm from the their<br />
apices.<br />
7. Verify the beam is in the center of M1 0. Adjust M5 if necessary, then<br />
repeat steps 5 and 6.<br />
8. Refer to "Optimizing <strong>Laser</strong> Output" earlier in this chapter to "walk<br />
the beam" and optimize laser output.<br />
9. Purge the system with clean, dry nitrogen gas when you are operating<br />
within water and oxygen absorption regions (Figure 7-1).<br />
10. Push the STATUS mode locker enable button (the ENABLE LED should<br />
be lit).<br />
11. Set up an autocorrelator such as the Model 409 -08 according to its<br />
user's manual, then maximize the signal by adjusting the angle of the<br />
nonlinear crystal for the optimum phase-matching condition.<br />
12. Maximize the intensity of the autocorrelated waveform by adjusting<br />
the coarse and fine PHASE controls.<br />
Refer to chapter 6, "Alignment: Coarse Phase Adjustment ...," for<br />
information on setting the coarse and fine PHASE controls.<br />
13. Adjust the GTI POSITION control to obtain a pulse.<br />
14. Optimize system performance by iterating between the following<br />
adjustments:<br />
a. Adjust the fine PHASE control.<br />
b. Adjust ion laser output power to obtain a stable pulse.<br />
c. Adjust the GTI POSITION control, prisms Pr2 and Pr3, the slit width,<br />
and pump source interactively to obtain the desired pulse width. I -"<br />
I