4<br />

<strong>Tsunami</strong><br />

Beam %<br />

2 mm (0.08 in.)<br />

Prism<br />

Figure 7-4: Placement of Pr, and Pr4 in beam<br />

5. Move prisms Prl and Pr4 SO that the beam passes through them about<br />

2 mm from their apices (Figure 7-4). For x-long wavelengths, ensure<br />

the beam is not clipped by these prisms.<br />

6. Use the prism dispersion compensation control to adjust Pr2 and Pr3<br />

so the beam also passes through them about 2 mm from the their<br />

apices.<br />

7. Verify the beam is in the center of M1 0. Adjust M5 if necessary, then<br />

repeat steps 5 and 6.<br />

8. Refer to "Optimizing <strong>Laser</strong> Output" earlier in this chapter to "walk<br />

the beam" and optimize laser output.<br />

9. Purge the system with clean, dry nitrogen gas when you are operating<br />

within water and oxygen absorption regions (Figure 7-1).<br />

10. Push the STATUS mode locker enable button (the ENABLE LED should<br />

be lit).<br />

11. Set up an autocorrelator such as the Model 409 -08 according to its<br />

user's manual, then maximize the signal by adjusting the angle of the<br />

nonlinear crystal for the optimum phase-matching condition.<br />

12. Maximize the intensity of the autocorrelated waveform by adjusting<br />

the coarse and fine PHASE controls.<br />

Refer to chapter 6, "Alignment: Coarse Phase Adjustment ...," for<br />

information on setting the coarse and fine PHASE controls.<br />

13. Adjust the GTI POSITION control to obtain a pulse.<br />

14. Optimize system performance by iterating between the following<br />

adjustments:<br />

a. Adjust the fine PHASE control.<br />

b. Adjust ion laser output power to obtain a stable pulse.<br />

c. Adjust the GTI POSITION control, prisms Pr2 and Pr3, the slit width,<br />

and pump source interactively to obtain the desired pulse width. I -"<br />