Alignment<br />

M5 fs Position M5 ps Position Input Bezel<br />

\ \ \<br />

Output Bezel M4 ps Position M4 fs Position<br />

Figure 6-8: Locations of M4 and M5 for fs and ps operation.<br />

b. Adjust M4 vertically and horizontally to center the reflected<br />

images from Mg and Mj onto Ms.<br />

If you have a ps-only system without prisms, skip to step f. If you are<br />

converting your <strong>Tsunami</strong> from a fs to a ps laser, or vice versa, refer to<br />

the section "Converting Between fs and ps Configurations" later in<br />

this chapter and perform the appropriate procedure, then return here.<br />

c. Close the shutter.<br />

d. Remove the slit assembly.<br />

WARNING<br />

Support the slit assembly with your hand when loosening the setscrew<br />

to prevent it from falling off the shaft and damaging the optics below.<br />

Loosen the setscrew holding the slit assembly to the spindle of<br />

the control micrometer, then slide the slit down and off the micrometer<br />

shaft (Figure 6-9).<br />

e. Open the shutter.<br />

f. Place a white card in front of the input to the A10 modulator<br />

(AOM).