Alignment<br />
WARNING<br />
The rotational alignment of the Ti:sapphire rod has been optimized at<br />
the factory. DO NOT remove the rod from its mount or alter its alignment,<br />
except to adjust its translation (as detailed below). Any other modification<br />
or adjustment, other than those made by a qualified Spectra-Physics<br />
<strong>Laser</strong>s service engineer, will void the warranty.<br />
c. Secure the rod locking screws.<br />
20. Maximize output power by optimizing the pump beam waist position<br />
in the rod.<br />
a. Close the shutter.<br />
b. Barely loosen the P2 setscrew located on the bottom of the mount<br />
(Figure 6-6).<br />
When loose, P2 should not wiggle when you try to move it by<br />
hand.<br />
c. Open the shutter.<br />
d. Set pump laser for nominal output power.<br />
e. Note <strong>Tsunami</strong> output power.<br />
f. Turn the P2 translation screw turn in either direction, and adjust<br />
P2 vertically and horizontally for maximum output power<br />
(Figure 6-6).<br />
g. Optimize Ml and M1 for maximum power.<br />
Since the P2 adjustment is coarse, you must also optimize Mlo and<br />
MI to see if power was gained or lost.<br />
If step f results in increased output power, continue repeating<br />
steps f and g until you reach maximum power. If the power level<br />
decreased, repeat these steps, but turn the translation screw in the<br />
opposite direction.<br />
h. Reduce pump power to minimum.<br />
i. Close the shutter.<br />
j. Tighten the P2 setscrew.<br />
k. Open the shutter.<br />
1. Increase pump laser to nominal output power.<br />
m. Adjust Pp horizontally and vertically for maximum output power.<br />
If unit stops lasing, readjust P2 vertically and horizontally to<br />
re-establish lasing and to maximize output power.<br />
Continue with "Aligning the Photodiode PC Board."